Noise (computed tomography)
Noise in computed tomography is an unwanted change in pixel values in an otherwise homogenous image. Often noise is defined loosely as the grainy appearance on cross-sectional imaging; more often than not, this is quantum mottle.
Noise in CT is measured via the signal to noise ratio (SNR); comparing the level of desired signal (photons) to the level of background noise (pixels deviating from normal). The higher the ratio, the less noise is present in the image.
Noise in a cross-sectional image will equal a decrease in the picture quality and inadvertently will hinder the contrast resolution.
Factors affecting noise
mAs
The mAs or the dose of a CT scan has a direct relationship with the number of photons utilized in the examination. A useful relationship to keep in mind is:
2 x mAs = 40% increase SNR
Increasing the dose of the scan will decrease the amount of noise and hence improve the contrast resolution of the image. However it comes at a cost, and balancing the dose with the contrast resolution required for interpretation must be considered when determining examination settings.
Studies that rely on superior contrast resolution will inescapably require a higher dose than examinations that can tolerate a higher amount of noise, for example, liver imaging vs. cardiac calcium scores.
Slice thickness
The number of photons available to generate an image has a linear relationship to the slice thickness. The thicker the slice, the more photons available; and the more photons available, the better the SNR. However, this isn't without a trade off because increasing the slice thickness will decrease the spatial resolution in the z-axis.
Patient size
Larger patients will absorb more radiation than smaller ones, meaning fewer photons will reach the detector hence reducing the signal to noise ratio.
Reconstruction algorithm
Nonlinear reconstruction algorithms can cause noise non-uniformity, meaning the intensity of noise varies across the image depending on regional structure. Uniform regions of the image will generally have lower noise levels than highly structured regions.
Noise metrics
A variety of metrics are used to measure different qualities of CT noise. Noise has many aspects including magnitude, texture, and nonuniformity.
Magnitude
Noise magnitude is quantified simply by the standard deviation. CT noise magnitude makes up the denominator of the signal to noise ratio.
Texture
Noise texture is the visual impression or quality of noise. It can be measured quantitatively by computing the noise power spectrum.
Nonuniformity
Noise nonuniformity is caused by variation in noise magnitude or texture across the image.
Related Radiopaedia articles
Imaging technology
- imaging technology
- imaging physics
- imaging in practice
-
x-ray
- x-ray physics
- x-ray in practice
- x-ray production
- x-ray tubes
- tube rating
- filters
- automatic exposure control (AEC)
- beam collimators
- grids
- air gap technique
- cassette
- intensifying screen
- x-ray film
- image intensifier
- digital radiography
- digital image
- mammography
- x-ray artifacts
- radiation units
- radiation safety
- radiation detectors
- fluoroscopy
-
computed tomography (CT)
- CT physics
- CT in practice
- CT technology
- CT image reconstruction
- CT image quality
- CT dose
- CT contrast media
-
CT artifacts
- patient-based artifacts
- physics-based artifacts
- hardware-based artifacts
- ring artifact
- tube arcing
- out of field artifact
- air bubble artifact
- helical and multichannel artifacts
- CT safety
- history of CT
-
MRI
- MRI physics
- MRI in practice
- MRI hardware
- signal processing
-
MRI pulse sequences (basics | abbreviations | parameters)
- T1 weighted image
- T2 weighted image
- proton density weighted image
- chemical exchange saturation transfer
- CSF flow studies
- diffusion weighted imaging (DWI)
- echo-planar pulse sequences
- fat-suppressed imaging sequences
- gradient echo sequences
- inversion recovery sequences
- metal artifact reduction sequence (MARS)
-
perfusion-weighted imaging
- techniques
- derived values
- saturation recovery sequences
- spin echo sequences
- spiral pulse sequences
- susceptibility-weighted imaging (SWI)
- T1 rho
- MR angiography (and venography)
-
MR spectroscopy (MRS)
- 2-hydroxyglutarate peak: resonates at 2.25 ppm
- alanine peak: resonates at 1.48 ppm
- choline peak: resonates at 3.2 ppm
- citrate peak: resonates at 2.6 ppm
- creatine peak: resonates at 3.0 ppm
- functional MRI (fMRI)
- gamma-aminobutyric acid (GABA) peak: resonates at 2.2-2.4 ppm
- glutamine-glutamate peak: resonates at 2.2-2.4 ppm
- Hunter's angle
- lactate peak: resonates at 1.3 ppm
- lipids peak: resonates at 1.3 ppm
- myoinositol peak: resonates at 3.5 ppm
- MR fingerprinting
- N-acetylaspartate (NAA) peak: resonates at 2.0 ppm
- propylene glycol peak: resonates at 1.13 ppm
-
MRI artifacts
- MRI hardware and room shielding
- MRI software
- patient and physiologic motion
- tissue heterogeneity and foreign bodies
- Fourier transform and Nyquist sampling theorem
- MRI contrast agents
- MRI safety
-
ultrasound
- ultrasound physics
-
transducers
- linear array
- convex array
- phased array
- frame averaging (frame persistence)
- ultrasound image resolution
- imaging modes and display
- pulse-echo imaging
- real-time imaging
-
Doppler imaging
- Doppler effect
- color Doppler
- power Doppler
- B flow
- color box
- Doppler angle
- pulse repetition frequency and scale
- wall filter
- color write priority
- packet size (dwell time)
- peak systolic velocity
- end-diastolic velocity
- resistive index
- pulsatility index
- Reynolds number
- panoramic imaging
- compound imaging
- harmonic imaging
- elastography
- scanning modes
- 2D ultrasound
- 3D ultrasound
- 4D ultrasound
- M-mode
-
ultrasound artifacts
- acoustic shadowing
- acoustic enhancement
- beam width artifact
- reverberation artifact
- ring down artifact
- mirror image artifact
- side lobe artifact
- speckle artifact
- speed displacement artifact
- refraction artifact
- multipath artifact
- anisotropy
- electrical interference artifact
- hardware-related artifacts
- Doppler artifacts
- aliasing
- tissue vibration
- spectral broadening
- blooming
- motion (flash) artifact
- twinkling artifact
- acoustic streaming
- biological effects of ultrasound
- history of ultrasound
-
nuclear medicine
- nuclear medicine physics
- detectors
- tissue to background ratio
-
radiopharmaceuticals
- fundamentals of radiopharmaceuticals
- radiopharmaceutical labeling
- radiopharmaceutical production
- nuclear reactor produced radionuclides
- cyclotron produced radionuclides
- radiation detection
- dosimetry
- specific agents
- carbon-11
- chromium-51
- fluorine agents
- gallium agents
- Ga-67 citrate
- Ga-68
- iodine agents
-
I-123
- I-123 iodide
- I-123 ioflupane (DaTSCAN)
- I-123 ortho-iodohippurate
- I-131
-
MIBG scans
- I-123 MIBG
- I-131 MIBG
-
I-123
- indium agents
- In-111 Octreoscan
- In-111 OncoScint
- In-111 Prostascint
- In-111 oxine labeled WBC
- krypton-81m
- nitrogen-13
- oxygen-15
- phosphorus-32
- selenium-75
-
technetium agents
- Tc-99m DMSA
- Tc-99m DTPA
- Tc-99m DTPA aerosol
- Tc-99m HMPAO
- Tc-99m HMPAO labeled WBC
- Tc-99m MAA
- Tc-99m MAG3
- Tc-99m MDP
- Tc-99m mercaptoacetyltriglycine
- Tc-99m pertechnetate
- Tc-99m labeled RBC
- Tc-99m sestamibi
- Tc-99m sulfur colloid
- Tc-99m sulfur colloid (oral)
- thallium-201 chloride
- xenon agents
- in vivo therapeutic agents
- pharmaceuticals used in nuclear medicine