Normal paper clip test of an ultrasound probe

Case contributed by Bálint Botz
Diagnosis not applicable

Presentation

Routine inspection.

A standard gem type metal paper clip is translated along the surface of the probe after application of a thin layer of US gel. No focal defects in the resultant reverberation artifact can be observed. 

Case Discussion

The paper clip test is a sensitive method to look for small piezo element dropouts. Note that due to transducer geometry the standard approach is not suitable for phased array probes, hence a modifed technique utilizing B-mode has been proposed. 