Time gain compensation

Changed by Felipe Campos Kitamura, 9 Aug 2014

Updates to Article Attributes

Body was changed:

Ultrasound imaging is based on returned ultrasound echoes from tissues. It is know that early echoes represent wave reflections in superficial layers, while late echoes come from deep layers. This is the base of axial localisation in ultrasound. 

It is also know that the emitted ultrasound wave amplitude gets smaller as it penetrates tissue, a phenomenon called attenuation. So, one might expect late echoes (from deep layers) to have smaller amplitudes than early (superficial) echoes even if those layers have the same echogenicity.

If the ultrasound image was formed directly by the raw returned echoes, image would appear lighter in superficial layers and darker in deep layers.

A way to overcome this problem is a time dependent gain compensation, in which signal gain is increased as time passes from the emitted wave pulse. This correction makes equally echogenic tissues look the sameeven if they are located in different depths.

  • -<p>Ultrasound imaging is based on returned ultrasound echoes from tissues. It is know that early echoes represent wave reflections in superficial layers, while late echoes come from deep layers. This is the base of axial localisation in ultrasound. </p><p>It is also know that the emitted ultrasound wave amplitude gets smaller as it penetrates tissue, a phenomenon called attenuation. So, one might expect late echoes (from deep layers) to have smaller amplitudes than early (superficial) echoes even if those layers have the same echogenicity.</p><p>If the ultrasound image was formed directly by the raw returned echoes, image would appear lighter in superficial layers and darker in deep layers.</p><p>A way to overcome this problem is a time dependent gain compensation, in which signal gain is increased as time passes from the emitted wave pulse. This correction makes equally echogenic tissues look the same even if they are located in different depths.</p>
  • +<p>Ultrasound imaging is based on <strong>returned ultrasound echoes</strong> from tissues. It is know that early echoes represent wave reflections in superficial layers, while late echoes come from deep layers. This is the base of axial localisation in ultrasound. </p><p>It is also know that the emitted ultrasound wave amplitude gets smaller as it penetrates tissue, a phenomenon called <strong>attenuation</strong>. So, one might expect late echoes (from deep layers) to have smaller amplitudes than early (superficial) echoes even if those layers have the same echogenicity.</p><p>If the ultrasound image was formed directly by the <strong>raw returned echoes</strong>, image would appear <strong>lighter in superficial layers and darker in deep layers</strong>.</p><p>A way to overcome this problem is a <strong>time dependent gain compensation</strong>, in which signal gain is increased as time passes from the emitted wave pulse. This correction makes <strong>equally echogenic tissues look the same</strong> <strong>even if they are located in different depths</strong>.</p>

Tags changed:

  • gain; time gain compensation; echo

Sections changed:

  • Imaging Technology

ADVERTISEMENT: Supporters see fewer/no ads

Updating… Please wait.

 Unable to process the form. Check for errors and try again.

 Thank you for updating your details.